All Categories
Dialyzer Reprocessing Machine

Home /  Products  /  Hemodialysis Product  /  Dialyzer Reprocessing Machine

Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine
Dialyzer For Hemodialysis machine

Dialyzer For Hemodialysis machine

A hemodialyzer is a key device used in hemodialysis treatment, whose main role is to purify the blood by removing waste and excess water from the blood through an artificial semi-permeable membrane. The working principle of the hemodialyzer is based on the action of solute diffusion, osmotic pressure difference and transmembrane pressure. The dialyzer consists of 8,000-12,000 hollow fibers with an inner diameter of about 200 microns.

Product Brochure:DOWNLOAD

  • Introduction
  • Parameter

Introduction

Dialyzer For Hemodialysis machine supplier

Parameter
Low-Flux Hollow Fiber Dialyzer
SM120L
SM140L
SM160L
SM180L
SM200L
UF Coefficient(mL/h·mmHg)
( QB=200mL/min;TMP=100mmHg)
12
14
16
18
20
Effective surface area (㎡)
1.2
1.4
1.6
1.8
2
Clearance in vitro
(QB=200mL/min,
QD=500mL/min
QF=10mL/min)
Urea
175
177
189
191
193
Creatinine
159
161
179
183
185
Phosphate
150
155
160
165
170
Vitamin B12
95
105
110
115
120
Clearance in vitro
(QB=300mL/min
QD=500mL/min
QF=10mL/min)
Urea
225
229
243
251
256
Creatinine
211
214
220
231
238
Phosphate
200
213
220
230
240
Vitamin B12
100
112
120
130
140
High-Flux Hollow Fiber Dialyzer
SM120H
SM140H
SM160H
SM180H
SM200H
UF Coefficient(mL/h·mmHg)
(QB=200ml/min;TMP=1000mmHg)
48
54
60
65
70
Effective surface area (㎡)
1.2
1.4
1.6
1.8
2
Sieving
coefficient
Inulin
0.9×(1±10%)
B 2-microglobulin
≥0.6
Myoglobin
≥0.50
Albumin
≤0.01
Clearance in vitro
(QB=200mL/min
QD=500mL/min,
QF=10mL/min)
Urea
191
193
195
197
198
Creatinine
181
183
185
190
195
Phosphate
176
178
181
185
190
Vitamin B12
135
145
155
165
175
Clearance in vitro
(QB=300mL/min,
OD=500mL/min,
QF=10mL/min)
Urea
255
260
267
275
280
Creatinine
230
240
250
260
270
Phosphate
215
225
235
250
262
Vitamin B12
140
157
175
195
208
Internal Fibre Diameter
200 μm
Membrane Thickness
40 μm
Membrane Type
Polyethersulfone
potting material
Polyurethane
Sterilization Method
Dry gamma