Medical Infusion Pump
Unique Rotary number setter
Dual-CPU design
Compatible with various brands of infusion sets
Comprehensive alarm function
Product Brochure:DOWNLOAD
- Introduction
- Parameter
Introduction
• Ultrasonic bubble detection technology
• Dynamic pressure detection technology
• Multiple infusion modes
• Emergency power supply
• Nursing call port RS232
• User-Friendly, Entry-level products preferred
Parameter
Technical specification:
Infusion Rate range (ml/h) |
1-1200 1.0ml/h-99.9ml/h range adjustable in 0.1ml/h steps or 1ml/h steps 100.0ml/h-1200ml/range adjustable in 1ml/h steps only |
Infusion precision |
±5%;( When correctly calibrated and operated in a stable environment) |
Pre-set of infusion volu |
0-9999 ml, 1ml steps |
Infusion volume accuracy |
±5%;( When correctly calibrated and operated in a stable environment) |
Bolus rate( ml/h) and accuracy |
200~1200, adjustable, default ±20%: |
Max pressure |
>120kPa |
Occlusion alarm level |
50KPa±25KPa |
KVO Rate |
KVO rate 3ml/h, flow rate> 10ml/h KVO rate 1ml/h, Iml/h <flow rate 10ml/h |
Recoverable alarm time |
1 min 50s *2 min |
Pause overtime alarm time |
1 min 50+ 2 min |
Bubble sensitivity |
Detectability of air bubble >5ouL |
Alarm |
Air in line, Occlusion, Door Open, Complete infusion, Low Power, Battery Exhausted, Motor Failure, Low Temperature, Pause Overtime etc |
Power |
ac.100-240V50/60Hz |
AC |
40VA |
Gross Weight |
2.5kg |
Classification |
Type CF, Class I, IPX1 |
Size |
165mm(L)×120mm(W)×220mm(H) |
OperationConditions |
Operating Temperature: +5℃-+40℃ Relative Humidity: 20%-90% |
Atmospheric pressure: 70KPa-106KPa |
|
Infusion pump should be operated in conditions free from strong shocks or vibrations, strong magnetic fields or corrosive gas environments. |
|
Environmental temperature for Charging battery |
Environmental Temperature:-20℃++60℃ |