POCT 4 in1 Multi-Monitoring Meter
Auto-Strip Recognition
AVTN Display
AAA batteries or portable power supply
Fast Test Time
Product Brochure:DOWNLOAD
- Introduction
- Parameter
Introduction
1.Measuring port- Insert test strip here to turn meter on for testing
2.C Button-Used to change date and time, code number and review test
results in memory
3. M Button- Used to turn meter on to enter setting mode and memory mode.
4.Display-Symbols, Simple message and test results appear here.
5.Ejection button- Press here to eject the used strip out.
Parameter
Specification
|
|
Power supply
|
AAA batteries or portable power supply;
|
Working environment
|
Humidity10%-90%,Temperature10-40℃;
|
Measuring speed
|
Total Cholesterol 15 Sec;Uric Acid 5 Sec;Blood Glucose 5 Sec;Hemoglobin 5 Sec.
|
Test Principle
|
Electrochemical Biosensor
|
Measuring range
|
Total Cholesterol:103-413mg/dL(2.58-10.34mmol/L):Uric Acid:1.5-19.83mg/dL(90-1190μmol/L):
Blood Glucose:20-600mg/dL(1.1-33.3mmol/L):Hemoglobin:5-26 g/dL |
Sample volume
|
Blood Glucose about 0.5pL blood,Total Cholesterol:about 5uL blood,
Uric Acid about 0.5uL blood,Hemoglobin about 0.5uL blood: |
Sample
|
Capillary whole blood and venous whole blood;
|
Unit of measurement
|
mg/dL,mmol/L.umol/L.g/dL
|
Storage temperature
|
-35℃-+55℃
|
Warranty
|
5 years
|
Weight
|
110g
|
Size
|
111mmx59mmx16mm
|
Hematocrit Range
|
Total Cholesterol 20-60%;Uric Acid 20-60%;Blood Glucose 20-60%
|
Precision
|
|
System Precision of Total Cholesterol
|
|
The average total cholesterol concentration
|
Precision
|
140mg/dL
|
SD=14mg/dL
|
220mg/dL
|
CV=4.13%
|
268 mg/dL
|
CV=4.25%
|
340mg/dL
|
CV=5.12%
|
384mg/dL
|
CV=5.65%
|
System Precision of Uric Acid
|
|
The average uric acid concentration
|
Precision
|
2.98mg/dL
|
SD=0.32mg/dL
|
5.3mg/dL
|
CV=4.13%
|
7.13mg/dL
|
CV=4.25%
|
13.08mg/dL
|
CV=5.12%
|
18.35mg/dL
|
CV=3.65%
|
System Precision of the Blood Glucose
|
|
The average blood glucose concentration
|
Precision
|
49mg/dL
|
SD=3mg/dL
|
83mg/dL
|
CV=3.10%
|
130mg/dL
|
CV=2.32%
|
238mg/dL
|
CV=1.76%
|
369 mg/dL
|
CV=1.99%
|
System Precision of Hemoglobin
|
|
Test range of hemoglobin
|
Precision
|
5g/dL-10g/dL
|
SD<±2g/dL
|
10g/dL-26g/dL
|
CV<±7.5%
|