* LCD color screen,Multiple languages
* Hydrating while feeding
* Auto-flush & pre-flush for anti- Occlusion function
* Unique rotary number setter
* Dual-MCU design

Product Brochure:DOWNLOAD

  • Introduction
  • Parameter

Introduction

* Comprehensive alarm function
* Ultrasonic bubble detection technology
* Sensitive detection of dynamic pressure
* 72 hours feeding history available
* A variety of feeding modes:Continuous mode,Intermittent mode,newborn mode
* Emergency power supply
* Exclusive thermostat (optional)
* Exclusive feeding sets:spike+flushing bag/feeding bag+flushing bag
* 360 adjustable quick-mount pole clamp

Parameter
Infusion preset range (ml)
Feeding volume: 0-9999, adjusted by 1 Water feeding volume: 0-999, adjusted by 1
Infusion(feeding or watering) accuracy
±10% or±0.2ml/h(choose the larger value)
Feeding speed setting range (ml/h)
1~800, adjusted by 1
Feeding speed accuracy
±10% or±0.2ml/h(choose the larger value)
Water feeding speed and accuracy (ml/h)
1000±10%
Flushing speed and accuracy (ml/h)
1000±10%
Single Flushing dose (ml)
25±10%
Perfusion speed and precision (ml/h)
1000±10%
Perfusion dose (ml)
10ml±10% nutrition and 25ml±10% water
The Maximum infusion pressure
≤0.1 MPa
The trigger alarm time at minimum flow rate
42min18s
The trigger alarm time at medium flow rate
1min30s
The amount of bolus produced by Medium speed up to the blocking alarm threshold
0.49ml
Blocking alarm threshold (pressure)
50KPa±10KPa
The recoverable time when alarm is automatically restored after the time is out
1min50s - 2min
Pause time when alarm timeout
1min50s - 2min
The heating temperature
When the feeding speed is greater than 20ml/h,15cm distance heater outlet is 25-38ºC
Bubble sensitivity
A bubble that can detect a single volume greater than 50uL
Intestinal feeder
Special intestinal feeding device
Alarm function
Tube occlusion, battery exhaustion, overtemperature
Prompt function
Tubing Bubble, Infusion Completion, Flush Completed, Battery Under voltage, Equipment Failure (ER Series), Pause Timeout, Network
Power On / Off, Feeding /watering Switching
Noise
Internal rechargeable lithium battery 11.1V;
battery life > 8 hoursafter and above.
Power supply and input
100-240V ~ 50 /60Hz ≤40VA
Internal battery
Internal rechargeable lithium battery 11.1V;
battery life > 8 hours
Output
≤20VA
Rated output voltage
DC15V
Total Weight
About 1.5kg
Rate
Class II BF type continuous operation capacity type nutrition pump with internal power supply;
Class IPX1 (anti-dripping device)
Dimensions
135mm (L) ×105mm (W) ×160m (mH)
Accessory
1 User's Manual and other document, 1 adapter, 1 set of intestinal feeding device and heater